RIGOL 2 Command System
2-538 DS2000 Programming Guide
2-538
:TRIGger:SPI:CLEVel
Syntax
:TRIGger:SPI:CLEVel <level>
:TRIGger:SPI:CLEVel?
Description
Set the trigger level of SCL in SPI trigger and the unit is the same with the current
amplitude unit.
Query the current trigger level of SCL in SPI trigger.
Parameter
Name Type Range Default
<level> Real
± 5 × VerticalScale from the screen center -
OFFSet
0
Note:
For the VerticalScale, refer to the
:CHANnel<n>:SCALe command.
For the OFFSet, refer to the
:CHANNel<n>:OFFSet command.
Return Format
The query returns the trigger level in scientific notation.
Example
:TRIGger:SPI:CLEVel 0.16
The query returns 1.600000e-01.