Rigol DS2000 Series Dokumentacja Strona 171

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 616
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 170
2 Command System RIGOL
DS2000 Programming Guide
2-153
:CURSor:MANual:XDELta?
Syntax
:CURSor:MANual:XDELta?
Description
Query the difference
X between the X values at cursor A and cursor B in manual
cursor measurement.
Explanation
For the horizontal position of cursor A, refer to the
:CURSor:MANual:CAX
command.
For the horizontal position of cursor B, refer to the
:CURSor:MANual:CBX command.
The unit is determined by the horizontal unit currently selected (refer to
the
:CURSor:MANual:TUNit command).
Return Format
The query returns the current difference
X in scientific notation.
Example
:CURSor:MANual:XDELta?
The query returns 9.120000e-06.
Przeglądanie stron 170
1 2 ... 166 167 168 169 170 171 172 173 174 175 176 ... 615 616

Komentarze do niniejszej Instrukcji

Brak uwag