RIGOL 2 Command System
2-134 DS2000 Programming Guide
2-134
:CHANnel<n>:SCALe
Syntax
:CHANnel<n>:SCALe <scale>
:CHANnel<n>:SCALe?
Description
Set the vertical scale of the waveform of CH1 or CH2.
Query the current vertical scale of the waveform of CH1 or CH2.
Parameter
Name Type Range Default
<n> Discrete {1|2} --
<scale> Real 500μV to 10V 1V
Note: the range of the vertical scale is related to the probe ratio currently set. For the setting of
the probe ratio, refer to the
:CHANnel<n>:PROBe command.
Return Format
The query returns the vertical scale in scientific notation.
Example
:CHANnel1:SCALe 1
The query returns 1.000000e+00.