RIGOL 2 Command System
2-452 DS2000 Programming Guide
2-452
:TRIGger:RUNT:ALEVel
Syntax
:TRIGger:RUNT:ALEVel <level>
:TRIGger:RUNT:ALEVel?
Description
Set the upper limit of the trigger level in runt trigger and the unit is the same with the
current amplitude unit.
Query the current upper limit of the trigger level in runt trigger.
Parameter
Name Type Range Default
<level> Real
± 5 × VerticalScale from the screen center -
OFFSet
0
Note:
For the VerticalScale, refer to the
:CHANnel<n>:SCALe command.
For the OFFSet, refer to the
:CHANNel<n>:OFFSet command.
Return Format
The query returns the upper limit of the trigger level in scientific notation.
Example
:TRIGger:RUNT:ALEVel 0.16
The query returns 1.600000e-01.