2 Command System RIGOL
DS2000 Programming Guide
2-259
:MEASure:SETup:MAX
Syntax
:MEASure:SETup:MAX <value>
:MEASure:SETup:MAX?
Description
Set the upper limit of threshold measurement and the unit is %.
Query the current upper limit of threshold measurement.
Parameter
Name Type Range Default
<value> Integer 7 to 95 90
Explanation
The threshold is the vertical level (in percentage) being measured in the analog
channel. Setting the threshold will affect all the time, delay and phase parameters.
You can also use the
:MEASure:SETup:TYPE command to select the threshold
measurement setting.
Return Format
The query returns an integer between 7 and 95.
Example
:MEASure:SETup:MAX 95
The query returns 95.