RIGOL 2 Command System
2-438 DS2000 Programming Guide
2-438
:TRIGger:EDGe:LEVel
Syntax
:TRIGger:EDGe:LEVel <level>
:TRIGger:EDGe:LEVel?
Description
Set the trigger level of edge trigger and the unit is the same with the current
amplitude unit.
Query the current trigger level of edge trigger.
Parameter
Name Type Range Default
<level> Real
± 5 × VerticalScale from the screen center -
OFFSet
0
Note:
For the VerticalScale, refer to the
:CHANnel<n>:SCALe command.
For the OFFSet, refer to the
:CHANNel<n>:OFFSet command.
Return Format
The query returns the trigger level in scientific notation.
Example
:TRIGger:EDGe:LEVel 0.16
The query returns 1.600000e-01.