Programming Guide for DM3058
3-25
:measure:capacitance{<range>|MIN|MAX|DEF}
Function:
The command sets the measuring range of capacitance.
Parameters:
Name
Type
Range of Values
Default
<range>
discrete
{0|1|2|3|4|5 }
2
Explanations:
The table 3-6 indicates the parameters and their ranges.
Table 3-6 The capacitance parameters and their ranges
Parameter
Range
0
2 nF
1
20 nF
2
200 nF
3
2 μF
4
20 μF
5
200 μF
MIN
2 nF
MAX
200 μF
DEF
200 nF
Example:
To set the measuring range of capacitance as 2 μF:
:measure: capacitance3
34. :measure:capacitance:range?
Command Format:
:measure:capacitance:range?
Function:
The command queries the measuring range of capacitance.
Return Format: