Rigol DM3058 Dokumentacja Strona 130

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 264
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 129
RIGOL
Programming Guide for DM3058
Function:
The command resets all the DC current measuring parameters and trigger
parameters as defaults, and sets DC current measuring function.
Parameters:
Name
Type
Range of Values
Default
<range>
Discrete
{ 200 µA | 2 mA | 20 mA
| 200 mA | 2 A | 10 A }
200 mA
<resolution>
Discrete
{MIN|MAX|DEF}
Rate: medium
Explanations:
<range> may be 200 μA (MIN), 2 mA, 20 mA, 200 mA, 2 A, 10 A (MAX), AUTO
(autorange) (unit is A). MAX = 10, MIN = 0. 000 2.
<resolution> (unit is A); in default setting, the rate is medium.
6. CONFigure:DIODe
Command Format:
CONFigure:DIODe
Function:
The command resets all the diode measuring parameters and trigger parameters
as defaults, and sets diode measuring function.
7. CONFigure:FREQuency
Command Format:
CONFigure:FREQuency
[{<range>|MIN|MAX|DEF} [,{<resolution>|MIN|MAX|DEF}]]
Function:
The command resets all the frequency measuring parameters and trigger
parameters as defaults, and sets frequency measuring function.