The command sets or queries the enable state of 2WR NULL measuring function.
Explanations:
[SENSe:]RESistance:NULL[:STATe]□{ON|OFF}: Enable or disable 2WR NULL
measuring function.
[SENSe:]RESistance:NULL[:STATe]?: Query the enable state of 2WR NULL
measuring function.
The command ―[SENSe:]RESistance:NULL:VALue‖ can be used to set NULL value.
The commands ―CONFigure:RESistance‖ and ―MEASure:RESistance?‖ can be
used to disable NULL function (OFF).
2-wire and 4-wire resistance function both are influenced by the command. All
the command of 2WR and 4WR will have influence over the setting of resistance.
Return Format:
The command returns the enable state of 2WR NULL measuring function.
37. [SENSe:]RESistance:NULL:VALue
Command Format:
[SENSe:]RESistance:NULL:VALue□{<value>|MIN|MAX}
[SENSe:]RESistance:NULL:VALue?□[{MIN|MAX}]
Function:
The command sets or queries the NULL value of 2WR measurement.
Parameters:
Explanations:
[SENSe:]RESistance:NULL:VALue□{<value>|MIN|MAX}: Set the NULL value of
2WR measurement. MIN is -120% max range of current function, MAX is 120%
max range of current function, DEF = 0.
[SENSe:]RESistance:NULL:VALue?□[{MIN|MAX}]: Query the NULL value of 2WR
measurement.
Before sending the command, do run the command ―[SENSe:]RESistance:NULL
[:STATe]□ON‖ to enable NULL function.