RIGOL
Programming Guide for DM3058
<range> may be 200 mV (MIN), 2 V, 20 V, 200 V, 750 V (MAX), AUTO (autorange)
(unit is V). MAX = 750, MIN = 0. 2.
<resolution> (unit is V); in default setting, the rate is medium.
Return Format:
The query returns the data in output buffer.
11. MEASure[:VOLTage][:DC]?
Command Format:
MEASure[:VOLTage][:DC]?
[{<range>|AUTO|MIN|MAX|DEF} [,{<resolution>|MIN|MAX|DEF}]]
Function:
The command resets all the DC voltage measuring parameters and trigger
parameters as defaults, and sets DC voltage measuring function, gets the
measuring results and sends them to output buffer of instrument.
Parameters:
Name
Type
Range of Values
Default
<range>
Discrete
{200 mV | 2 V | 20 V
| 200 V | 1000 V}
20 V
<resolution>
Discrete
{MIN|MAX|DEF}
Rate: medium
Explanations:
<range> may be 200 mV (MIN), 2 V, 20 V, 200 V, 1000 V (MAX), AUTO
(autorange) (unit is V). MAX = 1000, MIN = 0. 2.
<resolution> (unit is V); in default setting, the rate is medium.
Return Format:
The query returns the data in output buffer.
Przeglądanie stron 149