RIGOL
Programming Guide for DM3058
11. CONFigure[:VOLTage]:AC
Command Format:
CONFigure[:VOLTage]:AC□
[{<range>|AUTO|MIN|MAX|DEF} [,{<resolution>|MIN|MAX|DEF}]]
Function:
The command resets all the AC voltage measuring parameters and trigger
parameters as defaults, and sets AC voltage measuring function.
Parameters:
{ 200 mV | 2 V | 20 V
| 200 V | 750 V }
Explanations:
<range> may be 200 mV (MIN), 2 V, 20 V, 200 V, 750 V (MAX) AUTO (autorange)
(unit is V). MAX = 750, MIN = 0. 2.
<resolution> (unit is V); in default setting, the rate is medium.
12. CONFigure[:VOLTage][:DC]
Command Format:
CONFigure[:VOLTage][:DC]□
[{<range>|AUTO|MIN|MAX|DEF} [,{<resolution>|MIN|MAX|DEF}]]
Function:
The command resets all the DC voltage measuring parameters and trigger
parameters as defaults, and sets DC voltage measuring function.
Parameters:
{200 mV | 2 V | 20 V
| 200 V | 1000 V}
Explanations: