Rigol DM3058 Dokumentacja Strona 164

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 264
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 163
RIGOL
Programming Guide for DM3058
1. [SENSe:]FUNCtion[:ON]
Command Format:
[SENSe:]FUNCtion[:ON]"<function>"
[SENSe:]FUNCtion[:ON]?
Function:
The command selects or queries measuring function.
Parameters:
Name
Type
Range of Values
Default
<function>
Discrete
{ 'volt:dc' | 'volt:ac' | 'curr:ac'
| 'curr:dc' | 'freq' | 'per' | 'cap'
| 'res' | 'fres' | 'cont' | 'diod' }
Required
parameter
Explanations:
[SENSe:]FUNCtion[:ON] "<function>": Select the measuring function for
instrument, the parameter enclosed by single quotation marks (‗‘) is a character
string related to measuring function. And the default is inexsitent, <function>
must be set. See table 5-2.
[SENSe:]FUNCtion[:ON]?: Query current measuring function.
Table 5-2 The parameters and their meanings
Parameter
Meaning
Parameter
Meaning
'volt:dc'
DCV
'cap'
capacitance
'volt:ac'
ACV
'res'
2WR
'curr:ac'
ACI
'fres'
4WR
'curr:dc'
DCI
'cont'
Continuity
'freq'
Frequency
'diod'
Diode
'per'
Period
Return Format:
The query returns the parameter related to current measuring function.