RIGOL
Programming Guide for DM3058
Return Format:
The query returns the enable state of 2WR interval time measuring function.
35. [SENSe:]RESistance:NPLC
Command Format:
[SENSe:]RESistance:NPLC□{<PLCs>|MIN|MAX|DEF}
[SENSe:]RESistance:NPLC?□[{MIN|MAX}]
Function:
The command sets or queries the interval time of 2-wire resistance measurement,
input value is multiples of PLC.
Parameters:
{0.006|0.02|0.06|0.2|1|2|10|100}
Explanations:
[SENSe:]RESistance:NPLC□{<PLCs>|MIN|MAX|DEF}: Set the interval time of
2-wire resistance measurement, input value is multiples of PLC. <PLCs> may be
0.006, 0.02, 0.06, 0.2, 1, 2, 10 or 100, MIN = 0.006 PLC, MAX = 100 PLC.
[SENSe:]RESistance:NPLC?□[{MIN|MAX}]: Query the interval time of 2-wire
resistance measurement, return value is multiples of PLC.
2-wire and 4-wire resistance function both are influenced by the command. All
the command of 2WR and 4WR will have influence over the setting of resistance.
Return Format:
The query returns the interval time of 2-wire resistance measurement, return
value is multiples of PLC.
36. [SENSe:]RESistance:NULL[:STATe]
Command Format:
[SENSe:]RESistance:NULL[:STATe]□{ON|OFF}
[SENSe:]RESistance:NULL[:STATe]?
Function: