Rigol DM3058 Dokumentacja Strona 200

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 264
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 199
RIGOL
Programming Guide for DM3058
of period measurement. Unit is s, so 400 ms should be input in form of 0.4.
[SENSe:]PERiod:APERture?[{MIN|MAX}]: Query the interval time of period
measurement.
Return Format:
The query returns the interval time of period measurement.
58. [SENSe:]PERiod:NULL[:STATe]
Command Format:
[SENSe:]PERiod:NULL[:STATe]{ON|OFF}
[SENSe:]PERiod:NULL[:STATe]?
Function:
The command sets or queries the enable state of period NULL measuring
function.
Explanations:
[SENSe:]PERiod:NULL[:STATe] {ON|OFF}: Enable or disable period NULL
measuring function.
[SENSe:]PERiod:NULL[:STATe]?: Query the enable state of period NULL
measuring function.
The command [SENSe:]PERiod:NULL:VALue can be used to set NULL value.
The commands CONFigure:PERiod and MEASure:PERiod? can be used to
disable NULL function (OFF).
Return Format:
The command returns the enable state of period NULL measuring function.
59. [SENSe:]PERiod:NULL:VALue
Command Format:
[SENSe:]PERiod:NULL:VALue{<value>|MIN|MAX}
[SENSe:]PERiod:NULL:VALue?[{MIN|MAX}]
Function: