RIGOL
Programming Guide for DM3058
5-79
[SENSe:]FRESistance:RANGe[:UPPer]{<range>|MIN|MAX|DEF}
[SENSe:]FRESistance:RANGe[:UPPer]? [{MIN|MAX}]
Function:
The command sets or queries the range of 4WR measurement.
Parameters:
Name
Type
Range of Values
Default
<range>
Discrete
{200 Ω | 2 kΩ | 20 kΩ | 200
| 2 MΩ | 10 MΩ | 100 MΩ}
200
Explanations:
[SENSe:]FRESistance:RANGe[:UPPer] {<range>|MIN|MAX|DEF}: Set the
range of 4WR measurement. <range> may be 200 Ω, 2 , 20 , 200 , 2 ,
10 MΩ or 100 MΩ, MIN = 200 Ω, MAX = 100 MΩ. Unit is Ω, so 2 kΩ‖ should be
input in form of 2000.
[SENSe:]FRESistance:RANGe[:UPPer]?[{MIN|MAX}]: Query the range of 4WR
measurement.
2-wire and 4-wire resistance function both are influenced by the command. All
the command of 2WR and 4WR will have influence over the setting of resistance.
Return Format:
The query returns the range of 4WR measurement.
51. [SENSe:]FRESistance:RESolution
Command Format:
[SENSe:]FRESistance:RESolution{<resolution>|MIN|MAX|DEF}
[SENSe:]FRESistance:RESolution?[{MIN|MAX}]
Function:
The command sets or queries the resolution of 4WR measurement.
Parameters:
Name
Type
Range of Values
Default
<resolution>
Discrete
{MIN|MAX|DEF}
Rate: medium