Rigol DM3058 Dokumentacja Strona 241

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 264
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 240
RIGOL
Programming Guide for DM3058
7-3
The programming steps——Agilent Commands:
0 cmdsetagilent /*Select Agilent commands*/
1 *CLS /*Clear the instrument state*/
2 *IDN? /*Query equipment ID to confirm if the instrument works
normally*/
3 /*Return: RIGOL Technoligies, DM3058, DM3A020080808, 99.00.00.00.00.00*/
4 conf:volt:dcauto /*Set DC voltage measuring function, auto mode*/
5 calc:staton /*Enable math operation function*/
6 calc:funcaver /*Set statistic fucntion*/
7 calc:aver:max? /*Query the maximum measured DC voltage*/
8 /*Return: 5.000104e-02*/
9 calc:aver:coun? /*Query the measured times*/
10 /*Return: 63*/
11 calc:statOFF /*Exit math operation function*/
Remarks:
1. After execute the fourth command, the default measuring rate will change to
―Fast‖. The measuring rate and the resolution are relevant. The rate can be
changed by setting different resolutions. For example, sending conf:volt:dc
auto,max will set the measuring rate to slow.
2. Command line 7 returns the maximum measured DC voltage.
3. Command line 9 returns the measured times.
Przeglądanie stron 240
1 2 ... 236 237 238 239 240 241 242 243 244 245 246 ... 263 264

Komentarze do niniejszej Instrukcji

Brak uwag