RIGOL 2 Command System
2-62 MSO2000A/DS2000A Programming Guide
:CALCulate:FFT:HSPan
Syntax
:CALCulate:FFT:HSPan <span>
:CALCulate:FFT:HSPan?
Description
Set or query the horizontal scale of the FFT operation result. The unit is Hz/div.
Parameter
Name Type Range Default
<span> Discrete Refer to the Explanation 1.25MHz/div
Explanation
The FFT horizontal scale is related to the current FFT sample rate of the screen (Sa). It can be Sa/20,
Sa/40, Sa/100 and Sa/200.
The current FFT sample rate of the screen = the number of pixels per grid horizontally/the horizontal
timebase. The number of pixels per grid horizontally is related to the on/off status of the FFT
anti-aliasing function; it is 50 when the ainti-aliasing function is disabled and 100 when the
anti-aliasing function is enabled. For the horizontal timebase, refer to the
:TIMebase[:MAIN]:SCALe
command.
You can set the horizontal scale of FFT operation indirectly by setting the FFT horizontal coefficient
(:CALCulate:FFT:HSCale
).
Return Format
The query returns the horizontal scale of the FFT operation result in scientific notation.