RIGOL 2 Command System
2-320 MSO2000A/DS2000A Programming Guide
:TRIGger:SHOLd:TYPe
Syntax
:TRIGger:SHOLd:TYPe <type>
:TRIGger:SHOLd:TYPe?
Description
Set or query the hold type of setup/hold trigger.
Parameter
Name Type Range Default
<type> Discrete {SETup|HOLd|SETHOLd} SETup
Explanation
SETup: the oscilloscope triggers when the setup time is less than the setting value (refer to
the
:TRIGger:SHOLd:STIMe command).
HOLd: the oscilloscope triggers when the hold time is less than the setting value (refer to
the
:TRIGger:SHOLd:HTIMe command).
SETHOLd: the oscilloscope triggers when the setup time or hold time is less than the corresponding
setting value (refer to the
:TRIGger:SHOLd:STIMe and :TRIGger:SHOLd:HTIMe commands).
Return Format
The query returns SET, HOL or SETHOL.