RIGOL 2 Command System
2-60 MSO2000A/DS2000A Programming Guide
:CALCulate:FFT:HSCale
Syntax
:CALCulate:FFT:HSCale <hscale>
:CALCulate:FFT:HSCale?
Description
Set or query the horizontal coefficient of FFT operation. This command indirectly sets the horizontal scale
of FFT operation.
Parameter
Name Type Range Default
<hscale> Discrete {1|2|3|4} 2
Note: You can use the :CALCulate:FFT:HSPan command to set the horizontal scale of FFT operation directly.
Explanation
The FFT horizontal coefficient represents the relationship between the FFT horizontal scale (Hspan)
and the current FFT sample rate of the screen (Sa).
1: Hspan=Sa/20
2: Hspan=Sa/40
3: Hspan=Sa/100
4: Hspan=Sa/200
The current FFT sample rate of the screen=the number of pixels per grid horizontally/the horizontal
timebase. The number of pixels per grid horizontally is related to the on/off status of the
aunti-aliasing function; it is 50 when the anti-aliasing function is disabled and 100 when the
anti-aliasing function is enabled. For the horizontal timebase, refer to the
:TIMebase[:MAIN]:SCALe
command.
Return Format
The query returns 1, 2, 3 or 4.