RIGOL 2 Command System
2-228 MSO2000A/DS2000A Programming Guide
:SAVE:REFerence[:STARt]
Syntax
:SAVE:REFerence[:STARt] {<internal_loc>|<file_spec>}
Description
Store the reference waveform currently stored by the oscilloscope into the internal or external memory in
“*.ref” format.
Parameter
Name Type Range Default
<internal_loc> ASCII string C:\LocalREF0.ref to C:\LocalREF9.ref --
<file_spec> ASCII string Refer to the Explanation --
Explanation
Before using this command to store the reference waveform file, send the
:REFerence:SAVe
command to store the reference waveform of the current reference channel; otherwise, this
command is invalid.
In internal storage, if the specified location already contains a file, the original file will be overwritten.
In external storage, <file_spec> includes the external storage directory and the filename with suffix.
Its length cannot exceeds 64 characters. Wherein, the storage directory can only be D:\. If the
specified directory already contains a file with the same filename, the original file will be overwritten.
Example
:SAVE:REFerence C:\LocalREF3.ref /*Store the reference waveform currently stored by the oscilloscope
into the internal memory*/