RIGOL Chapter 2 Command System
2-42 DS1000Z Programming Guide
:MATH:OPTion:END
Syntax :MATH:OPTion:END <end>
:MATH:OPTion:END?
Description
Set or query the end point of the waveform math operation.
Parameter
Name
Type
Range
Default
<end>
Discrete
(Current point currently set+1)~1199
1199
Explanation This command is invalid for FFT operation. Sending this command will modify the
end point of all operations (except FFT).
The horizontal axis of the source selected is equally divided into 1200 parts, in which
the left is numbered as 0 and the right is numbered as 1199.
Return
Format
The query returns an integer.
Example
:MATH:OPTion:END 200
Related
Command
:MATH:OPTion:STARt
:MATH:OPTion:INVert
Syntax :MATH:OPTion:INVert <bool>
:MATH:OPTion:INVert?
Description Enable or disable the inverted display mode of the operation result, or query the inverted
display mode status of the operation result.
Parameter
Name
Type
Range
Default
<bool>
Bool
{{0|OFF}|{1|ON}}
0|OFF
Explanation This command is invalid for FFT operation. Sending this command will modify the
inverted display mode status of all operations (except FFT).
This command has the same function as the :MATH:INVert command.
Return
Format
The query returns 0 or 1.
Example
:MATH:OPTion:INVert ON
Related
Command
:MATH:INVert