Rigol DS1000Z Series Dokumentacja Strona 22

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 172
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 21
RIGOL Chapter 2 Command System
2-8 DS1000Z Programming Guide
:CHANnel<n> Commands
:CHANnel<n> commands are used to set or query the vertical system parameters, such as bandwidth limit,
coupling, vertical scale and vertical offset.
Command List:
:CHANnel<n>:BWLimit
:CHANnel<n>:COUPling
:CHANnel<n>:DISPlay
:CHANnel<n>:INVert
:CHANnel<n>:OFFSet
:CHANnel<n>:RANGe
:CHANnel<n>:TCAL
:CHANnel<n>:SCALe
:CHANnel<n>:PROBe
:CHANnel<n>:UNITs
:CHANnel<n>:VERNier
:CHANnel<n>:BWLimit
Syntax :CHANnel<n>:BWLimit <type>
:CHANnel<n>:BWLimit?
Description
Set or query the bandwidth limit parameter of the specified channel.
Parameter
Name
Type
Range
Default
<n>
Discrete
{1|2|3|4}
——
<type>
Discrete
{20M|OFF}
OFF
Explanation OFF: disable the bandwidth limit and the high frequency components of the signal
under test can pass the channel.
20M: enable the bandwidth limit and the high frequency components that exceed 20
MHz are attenuated.
Enabling the bandwidth limit can reduce the noise, but also can attenuate the high
frequency components.
Return
Format
The query returns 20M or OFF.
Example
:CHANnel1:BWLimit 20M /*Enable the 20MHz bandwidth limit*/
Przeglądanie stron 21