Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-121
:TRIGger:RS232:PARity
Syntax :TRIGger:RS232:PARity <parity>
:TRIGger:RS232:PARity?
Description
Set or query the parity type when trigger condition is ERRor or PARity in RS232 trigger.
Parameter
Name
Type
Range
Default
<parity>
Discrete
{EVEN|ODD|NONE}
NONE
Explanation When the trigger condition is PARity, the parity type can not be set to NONE. At this
point, the default parity type is EVEN.
Return Format
The query returns EVEN, ODD or NONE.
Example
:TRIGger:RS232:PARity EVEN
Related
Command
:TRIGger:RS232:WHEN
:TRIGger:RS232:STOP
Syntax :TRIGger:RS232:STOP <bit>
:TRIGger:RS232:STOP?
Description
Set or query the stop bit when the trigger condition is ERRor in RS232 trigger.
Parameter
Name
Type
Range
Default
<bit> Discrete {1|2} 1
Return Format
The query returns 1 or 2.
Example
:TRIGger:RS232:STOP 2
Related
Command
:TRIGger:RS232:WHEN
:TRIGger:RS232:DATA
Syntax :TRIGger:RS232:DATA <data>
:TRIGger:RS232:DATA?
Description
Set or query the data when the trigger condition is DATA in RS232 trigger.
Parameter
Name
Type
Range
Default
<data>
Integer
0 to 2
n
-1
90
Explanation
In 2
n
-1, n is the current data bits and can be 5, 6, 7 or 8.
Return Format
The query returns an integer.
Example
:TRIGger:RS232:DATA 10
Related
Command
:TRIGger:RS232:WIDTh
:TRIGger:RS232:WHEN