Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-5
:ACQuire:MDEPth
Syntax :ACQuire:MDEPth <mdep>
:ACQuire:MDEPth?
Description Set or query the memory depth of the oscilloscope namely the number of waveform
points that can be stored in a single trigger sample. The default unit is pts (points).
Parameter
Name
Type
Range
Default
<mdep>
Discrete Refer to the Explanation
AUTO
Explanation When a single channel is on: {AUTO|12000|120000|1200000|12000000|24000000}
When dual channels are on: {AUTO|6000|60000|600000|6000000|12000000}
When four channels are on: {AUTO|3000|30000|300000|3000000|6000000}
Wherein, 24000000, 12000000 and 6000000 are optional.
The relationship among the memory depth, sample rate and waveform length is
satisfied with the following equation:
Memory Depth Sample Rate × Waveform Length
Wherein, the Memory Depth can be set using the
:ACQuire:MDEPth command and
the Waveform Length is the product of the horizontal timebase (set by
the
:TIMebase[:MAIN]:SCALe command) and the number of the horizontal scales
(12 for DS1000Z).
When AUTO is selected, the oscilloscope will select the memory depth automatically
according to the current sample rate.
Return
Format
The query returns the actual number of points (integer) or AUTO.
Example
:ACQuire:MDEPth 12000 /*Set the memory depth to 12000*/
:ACQuire:MDEPth? /*The query returns 12000*/
Related
Command
:ACQuire:SRATe?
:TIMebase[:MAIN]:SCALe