Programming Guide for DM3058/DM3058E
3-46
1. :RATE:VOLTage:DC
Command Format:
:RATE:VOLTage:DC?
:RATE:VOLTage:DC {F|M|S}
Function:
The commands query and set the rate of DC voltage measurement.
Explanations:
:RATE:VOLTage:DC? - Queries the current measuring rate of DC voltage. The
command is valid only when DCV function is enabled.
:RATE:VOLTage:DC {F|M|S} Sets the desired measuring rate of DC voltage. For
the range and the explanation of each parameter, see table below.
Table 3-10 The parameter range and explanations
Parameter Explanation Mark Rate Refresh rate
F (FAST) Fast measure F 123 reading/s 50 Hz
M (MEDIUM) Medium measure M 20 reading/s 20 Hz
S (SLOW) Slow measure S 2.5 reading/s 2.5 Hz
Return Format:
The query returns the parameter related to current rate (see Table 3-10).
Example:
Sets the rate under DC voltage measurement to Medium.
:RATE:VOLTage:DC M
2. :RATE:VOLTage:AC
Command Format:
:RATE:VOLTage:AC?
:RATE:VOLTage:AC {F|M|S}
Function:
The commands query and set the rate of AC voltage measurement.
Explanations: