RIGOL
Programming Guide for DM3058/DM3058E
3-37
MAX 100
MIN 200 Ω
DEF 200 kΩ
Example:
Sets the 2-wire resistance range to 200Ω.
:MEASure:RESistance MIN or :MEASure:RESistance 0
20. :MEASure:RESistance:RANGe?
Command Format:
:MEASure:RESistance:RANGe?
Function:
The command queries the current 2-wire resistance range.
Return Format:
The query returns: 0, 1, 2, 3, 4, 5 or 6 (see Table 3-8).
21. :MEASure:FRESistance?
Command Format:
:MEASure:FRESistance?
Function:
The command queries the measured 4-wire resistance value.
Explanations:
The command is unavailable in double-display.
Return Format:
The command queries the measured 4-wire resistance value in scientific notation,
such as: 8.822946e-05, the unit is Ω.
22. :MEASure:FRESistance
Command Format:
Przeglądanie stron 61