Programming Guide for DM3058/DM3058E
3-39
Return Format:
The command queries the measured frequency value of AC signal in scientific
notation, such as: 8.485240e-05, the unit is Hz.
25. :MEASure:FREQuency
Command Format:
:MEASure:FREQuency {<range>|MIN|MAX|DEF}
Function:
The command sets the voltage range of the input signal under frequency
measurement.
Parameter:
Name Type Range Default
<range> Discrete {0|1|2|3|4} 2
Explanations:
The available frequency range is 20Hz~1MHz.
For the parameter range, refer to Table 3-5.
Example:
Sets the voltage range under frequency measurement to 20 V.
:MEASure:FREQuency DEF
26. :MEASure:FREQuency:RANGe?
Command Format:
:MEASure:FREQuency:RANGe?
Function:
The command queries the AC voltage range used under frequency measurement.
Explanations:
The range here is the same with the range of AC voltage, see Table 3-5.