Rigol DM3058 Instrukcja Użytkownika Strona 119

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 192
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 118
RIGOL
Programming Guide for DM3058/DM3058E
4-26
6. MEASure:FRESistance?
Command Format:
MEASure:FRESistance?
{<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Function:
The command presets the multimeter as specified range and resolution and
executes 4-wire resistance measurement, and then sends the readings to the
output buffer. For the parameter range and other related information, refer to
CONFigure:FRESistance.
Return Format:
The query returns the data stored in output buffer.
7. MEASure:FREQuency?
Command Format:
MEASure:FREQuency?
{<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Function:
The command presets the multimeter as specified range and resolution and
executes frequency measurement, and then sends the readings to the output
buffer. For the parameter range and other related information, refer to
CONFigure:FREQuency.
Return Format:
The query returns the data stored in output buffer.
8. MEASure:PERiod?
Command Format:
MEASure:PERiod? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Function:
Presets the multimeter as specified range and resolution and executes period
Przeglądanie stron 118
1 2 ... 114 115 116 117 118 119 120 121 122 123 124 ... 191 192

Komentarze do niniejszej Instrukcji

Brak uwag