RIGOL
Programming Guide for DM3058/DM3058E
3-6
Turn off the statistic function.
:CALCulate:STATistic:STATe OFF
7. :CALCulate:REL:OFFSet
Command Format:
:CALCulate:REL:OFFSet?
:CALCulate:REL:OFFSet {<range>|MIN|MAX|DEF|CURR}
Function:
The commands query and set the offset of REL operation.
Parameter:
Name Type Range Default
<range>
Consecutive
real number
See Table 3-2 0
Explanations:
:CALCulate:REL:OFFSet? – Queries the REL offset of current measurement.
:CALCulate:REL:OFFSet {<range>|MIN|MAX|DEF|CURR} – Sets the desired REL
offset for current measurement.
The value of <range> varies with measurement type and range. See table below,
every parameter could be a value with sixth decimals. All the set value is only
valid for the current measurement and the value should be less than the upper
limit of current range.
Table 3-2 Offset range of REL operation
Measurement
items
Range DEF Unit
DC Voltage ±1200 0 V
AC Voltage ±900 0 V
DC Current ±12 0 A
AC Current ±12 0 A
Resistance ±1.2e+08 0 Ω
Capacitance ±1.2e-02 0 F
Frequency ±1.2e+06 0 Hz