RIGOL
Programming Guide for DS1000B Series
2-68
In NoRMAL mode, different types of instruments have different sweep
ranges:
DS1204B instrument<scale_val> range1ns/div~50s/div.
DS1104B instrument<scale_val>range2ns/div~50s/div.
DS1064B instrument<scale_val>range5ns/div~50s/div.
In SCAN mode, <scale_val>range: 50ms ~ 50s.
Query Format:
:TRIGger:ALTernation:TimeSCALe?[<source>]
Return Format:
The query returns the value of time scale, and the unit is s.
Example:
:TRIGger:ALTernation:TimeSCALe0.001,SOURCeB
Set the time scale as 1ms
:TRIGger:ALTernation:TimeSCALe?SOURCeB Return 1.000e-03.
(5) :TRIGger:ALTernation:TimeOFFSet
Command Format:
:TRIGger:ALTernation:TimeOFFSet<value>[,<source>]
Function:
The command sets the timebase offset.
In NORMAL mode, <value>: 1s ~ memory capacitance;
In STOP mode, <value: -500s ~ +500s;
Query Format:
:TRIGger:ALTernation:TimeOFFSet?[<source>]
Return Format:
The query returns the value of timebase offset, and the unit is s.
Example:
:TRIGger:ALTernation:TimeOFFSet0.0002,SOURceB
Set the timebase offset as 200μs.
:TRIGger:ALTernation:TimeOFFSet?SOURceB Return 2.000e-04.