RIGOL
Programming Guide for DS1000B Series
2-58
oscilloscope will not trigger until Holdoff ends.
The <count> range is 100ns~1.5s.
Query Format:
:TRIGger:HOLDoff?
Return Format:
The query returns the value of holdoff time, and the unit is s.
Example:
:TRIGger:HOLDoff0.0001 Set the holdoff time as 100μs.
:TRIGger:HOLDoff? Return 1.000e-04.
7. :TRIGger:STATus
Query Format:
:TRIGger:STATus?
Function:
The command queries the current status of the oscilloscope. The status may be
RUN, STOP, T’ D, WAIT,SCAN or AUTO.
Return Format:
The query returns RUN or STOP, T’D, WAIT, AUTO.
8. :TRIGger:SENSitivity
Command Format:
:TRIGger:SENSitivity<count>
Function:
The command sets the trigger sensitivity. The <count> range is 0.1div~1div.
Query Format:
:TRIGger:SENSitivity?
Return Format:
The query returns the value of trigger sensitivity, and the unit is div.