Rigol DS1204B Dokumentacja Strona 35

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 132
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 34
RIGOL
Programming Guide for DS1000B Series
2-27
Function:
The command measures the fall time of signal from <source>. The <source>
may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Return Format:
The query returns 4.000e-005, and the unit is s.
15. :MEASure:PERiod?
Command Format:
:MEASure:PERiod?[<source>]
Function:
The command measures the period of signal from <source>. The <source>
may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Return Format:
The query returns 1.000e-003, and the unit is s.
16. :MEASure:PWIDth?
Command Format:
:MEASure:PWIDth?[<source>]
Function:
The command measures the positive pulse width of signal from <source>. The
<source> may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Return Format:
The query returns 5.000e-004, and the unit is s.
17. :MEASure:NWIDth?
Command Format:
:MEASure:NWIDth?[<source>]
Function:
The command measures the negative pulse width of signal from <source>. The
<source> may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Przeglądanie stron 34
1 2 ... 30 31 32 33 34 35 36 37 38 39 40 ... 131 132

Komentarze do niniejszej Instrukcji

Brak uwag