Rigol DM3058 Dokumentacja Strona 250

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 264
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 249
RIGOL
Programming Guide for DM3058
Example 6: Reading Hold
Reading Hold function can get a stable reading and hold it on the screen even when
you remove the test leads.
The programming steps——RIGOL Commands:
0 *RST /*Reset the instrument for preparing measurement*/
1 cmdsetrigol / *Select RIGOL commands*/
2 *IDN? /*Query equipment ID to confirm if the instrument
works normally*/
3 /*Connect the test leads under DC voltage measurement mode*/
4 :function:voltage:DC /*Enable DC voltage measuring function*/
5 :measure:voltage:DC<range> /*Set the measurement range*/
6 :trigger:auto:holdon /*Enable reading hold function*/
7 :trigger:auto:hold:sensitivity1 /*Set the holding range as 0.1%*/
Remarks:
When setting the holding range in line 7 command , multiple thresholds can be set to
meet different requirements (refers to the corresponding commands in chapter 3:
Trigger subsystem).