5-77
47. [SENSe:]FRESistance:NULL:VALue
Command Format:
[SENSe:]FRESistance:NULL:VALue{<value>|MIN|MAX}
[SENSe:]FRESistance:NULL:VALue?[{MIN|MAX}]
Function:
The command sets or queries the NULL value of 4WR measurement.
Parameters:
Name
Type
Range of Values
Default
<value>
Discrete
± 120% max range
0
Explanations:
[SENSe:]FRESistance:NULL:VALue{<value>|MIN|MAX}: Set the NULL value of
4WR measurement. MIN is -120% max range of current function, MAX is 120%
max range of current function, DEF = 0.
[SENSe:]FRESistance:NULL:VALue? [{MIN|MAX}]: Query the NULL value of
4WR measurement.
Before sending the command, do run [SENSe:]FRESistance:NULL[:STATe]ON
to enable NULL function.
2-wire and 4-wire resistance function both are influenced by the command. All
the command of 2WR and 4WR will have influence over the setting of resistance.
Return Format:
The query returns the NULL value of 4WR measurement.
48. [SENSe:]FRESistance:OCOMpensated
Command Format:
[SENSe:]FRESistance:OCOMpensated{ON|OFF}
[SENSe:]FRESistance:OCOMpensated?
Function:
The command sets or queries the enable state of 4WR offset compensated
function.
Explanations: