Programming Guide for DM3058
5-31
The command resets all the frequency measuring parameters and trigger
parameters as defaults, and sets frequency measuring function, gets the
measuring results and sends them to output buffer of instrument.
Parameters:
Name
Type
Range of Values
Default
<range>
Consecutive
Integer
20 Hz~1 MHz
20 Hz
<resolution>
Discrete
{MIN|MAX|DEF}
Rate: medium
Explanations:
<range> may be 20 Hz~1 MHz (unit is Hz). MAX = 1 000 000, MIN = 20, DEF =
20.
<resolution> unit is Hz; in default setting, the rate is medium.
Return Format:
The query returns the data in output buffer.
7. MEASure:FRESistance?
Command Format:
MEASure:FRESistance?
[{<range>|AUTO|MIN|MAX|DEF} [,{<resolution>|MIN|MAX|DEF}]]
Function:
The command resets all the 4-wire resistance measuring parameters and trigger
parameters as defaults, and sets 4-wire resistance measuring function, get the
measuring results and send them to output buffer of instrument.
Parameters:
Name
Type
Range of Values
Default
<range>
Discrete
{200 Ω | 2 kΩ | 20 kΩ | 200 kΩ
| 2 MΩ | 10 MΩ | 100 MΩ}
2
<resolution>
Discrete
{MIN|MAX|DEF}
Rate: medium
Explanations:
<range> may be 200 Ω (MIN), 2 kΩ, 20 kΩ, 200 kΩ, 2 , 10 MΩ, 100 MΩ