Rigol DM3058 Dokumentacja Strona 173

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 264
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 172
RIGOL
Programming Guide for DM3058
5-57
Parameters:
Name
Type
Range of Values
Default
<resolution>
Discrete
{MIN|MAX|DEF}
Rate: medium
Explanations:
[SENSe:]CURRent[:DC]:RESolution {<resolution>|MIN|MAX|DEF}: Set the
resolution of DCI measurement. In default setting of <resolution>, the rate is
medium.
[SENSe:]CURRent[:DC]:RESolution?[{MIN|MAX}]: Query the resolution of DCI
measurement.
Return Format:
The query returns current resolution of DCI measurement.
16. [SENSe:]CURRent[:DC]:ZERO:AUTO
Command Format:
[SENSe:]CURRent[:DC]:ZERO:AUTO{ON|OFF}
[SENSe:]CURRent[:DC]:ZERO:AUTO?
Function:
The command sets or queries the enable state of DCI autozero function.
Explanations:
[SENSe:]CURRent[:DC]:ZERO:AUTO{ON|OFF}: Enable or disable DCI autozero
function.
[SENSe:]CURRent[:DC]:ZERO:AUTO?: Query the enable state of DCI autozero
function.
Return Format:
The query returns the enable state of DCI autozero function.
17. [SENSe:]VOLTage:AC:NULL[:STATe]
Command Format:
[SENSe:]VOLTage:AC:NULL[:STATe]{ON|OFF}
[SENSe:]VOLTage:AC:NULL[:STATe]?
Przeglądanie stron 172
1 2 ... 168 169 170 171 172 173 174 175 176 177 178 ... 263 264

Komentarze do niniejszej Instrukcji

Brak uwag