RIGOL Command Systems
Programming Guide for DS1000E, DS1000D Series
1. :WAVeform:DATA?
Command Format:
:WAVeform:DATA? [<source>]
Function Explanation:
The command reads the waveform data in the specified source. <source> could
be CHANnel1, CHANnel2, DIGital, MATH or FFT.
Returned Format:
The query returns 1024 data.
2. :WAVeform:POINts:MODE
Command Format:
:WAVeform:POINts:MODE <points_mode>
:WAVeform:POINts:MODE?
Function Explanation:
This command sets the mode of waveform points. <points_mode> can be:
NORMal, MAXimum or RAW.
Data points returned by :WAVeform:DATA? in different modes:
Normal Memory Long Memory In
state,
MAX is the
same with
NORMal; in
STOP state,
MAX is the
same with
RAW.
600 600 600
512 512 512
600 8192(8k) 524288(512k)
Channel
CHx*
600 16384(16k) 1048576(1M)
DIGITAL 600 16384(16k) 1048576(1M)
*Note: Half-channel means only one channel is open and the MATH is closed.
Returned Format:
The query returns NORMal, MAXimum or RAW.
Komentarze do niniejszej Instrukcji