RIGOL Command Systems
Programming Guide for DS1000E, DS1000D Series
2-8
:ACQuire:AVERages <count>
:ACQuire:AVERages?
Function Explanation:
The commands set and query the average numbers in Average mode. <count>
could be and integer of 2 times the power of N within 2 and 256.
Returned Format:
The query returns 2, 4, 8, 16, 32, 64, 128 or 256.
Example:
:ACQ:AVER 16 Setup the average acquisition time as 16.
:ACQ:AVER? The query returns 16.
4. :ACQuire:SAMPlingrate?
Command Format:
:ACQuire:SAMPlingrate? {CHANnel<n>|DIGITAL}
Function Explanation:
The command queries the current sampling rate of the analog channel or digital
channel (only for DS1000D series). <n> is 1 or 2 means channel 1 or channel 2.
Returned Format:
The query returns the setting value of the sampling rate.
Example:
:ACQ:SAMP? CHANnel2 Query the sampling rate for channel 2.
100000000.000000 The query returns 100M.
5. :ACQuire:MEMDepth <depth>
Command Format:
:ACQuire:MEMDepth <depth>
:ACQuire:MEMDepth?
Function Explanation: