Rigol DS1000D Instrukcja Użytkownika Strona 29

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 137
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 28
Command Systems RIGOL
Programming Guide for DS1000E, DS1000D Series
2-17
:TIM:OFFS? The query returns 1.000e+00.
3. :TIMebase[:DELayed]:SCALe
Command Format:
:TIMebase[:DELayed]:SCALe <scale_val>
:TIMebase[:DELayed]:SCALe?
Function Explanation:
The commands set and query the horizontal scale for MAIN or DELayed
timebase, the unit is s/div (seconds/grid), thereinto:
In YT mode, the range of <scale_val> is 2ns - 50s;
In ROLL mode, the range of <scale_val> is 500ms - 50s;
In MAIN state, the item [:DELayed] should be omitted.
Returned Format:
The query returns the setting value of <scale_val> in s.
Example:
:TIM:MODE MAIN Setup the timebase as MAIN.
:TIM:SCAL 2 Setup its scale as 2s.
:TIM:SCAL? The query returns 2.000e+00.
4. :TIMebase:FORMat
Command Format:
:TIMebase:FORMat <value>
:TIMebase:FORMat?
Function Explanation:
The commands set and query the horizontal timebase. <value> could be XY, YT
or SCANning.
Returned Format:
The query returns X-Y, Y-T or SCANNING.
Example:
Przeglądanie stron 28
1 2 ... 24 25 26 27 28 29 30 31 32 33 34 ... 136 137

Komentarze do niniejszej Instrukcji

Brak uwag