Programming Guide for DS1000/DS1000CA Series
6.6 :TRIGger:ALTernation:EDGE:SLOPe
Command Format:
:TRIGger:ALTernation:EDGE:SLOPe<value>
Explanation:
The command is to set the edge polarity of EDGE and SLOPE trigger in current
channel.
<value> can be POSitive or NEGative. The double quotes are not returned.
Query Format:
:TRIGger:ALTernation:EDGE:SLOPe?
Returned Format:
Query returnsPOSITIVE or NEGATIVE.
Example:
:TRIGger:ALTernation:EDGE:SLOPePOSitive
Set the edge polarity to Positive.
:TRIGger:ALTernation:EDGE:SLOPe?
Query returns POSITIVE.
6.7 :TRIGger:ALTernation<mode>:MODE
Command Format:
:TRIGger:ALTernation<mode>:MODE<value>
Explanation:
The command is to set the trigger modes and the corresponding conditions.
<mode> can be :PULSe, :VIDEO or :SLOPe.
When <mode> is :PULSe or :SLOPe”,
the options of <value> are +GREaterthan, +LESSthan, +EQUal,
-GREaterthan, -LESSthan or -EQUal.
When <mode> is :VIDEO”,
the options of <value> are ODDfield, EVENfield, LINE or ALLlines.
Query Format: