Rigol DS1000MD Instrukcja Użytkownika Strona 49

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 97
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 48
RIGOL
© 2008 RIGOL Technologies, Inc.
Programming Guide for DS1000/DS1000CA Series
7. :CHANnel<n>:SCALe
Command Format:
:CHANnel<n>:SCALe<range>
Explanation:
The command is to set the vertical range of the amplified waveform.
<range> is 2mV ~ 5V Probe 1X
10mV ~10V Probe 5X
20mV ~50v Probe 10X
100mV ~ 100V Probe 50X
200mV ~ 500v Probe 100X
1V ~1000V Probe 500X
2V ~5000V Probe 1000X
The options of n are 1 or 2.
Query: Format:
:CHANnel<n>:SCALe?
Returned Format:
Query returns the set value of <range> in scientific numeric notation.
Example:
:CHANnel2:PROBe10 Set probe attenuation as 10X
:CHANnel2:SCALe20 Set vertical scale as 20V
:CHANnel2:SCALe? Query returns 2.000e+01
8. :CHANnel<n>:FILTer
Command Format:
:CHANnel<n>:FILTer{ON|OFF}
Explanation:
The command is to set the digital filter ON or OFF. The options of n are 1 or 2.
Query Format:
:CHANnel<n>:FilTer?
Przeglądanie stron 48
1 2 ... 44 45 46 47 48 49 50 51 52 53 54 ... 96 97

Komentarze do niniejszej Instrukcji

Brak uwag