2-63
30~2000ms, and the default is 30ms; the resolution is , 200~2000ms, and the
default is 200ms; the resolution is 5½ , 400~2000ms, and the default is 400ms.
3. :trigger:auto:hold
Syntax:
:trigger:auto:hold?
:trigger:auto:hold
<value>
Function:
The command queries and sets auto trigger delay state.
Parameters:
Name
Type
Range
Default
<value>
Boolean
{ON|OFF}
OFF
Return Format:
:trigger:auto:hold? returns the current auto trigger delay state. ON means the delay
is on, and OFFmeans off.
4. :trigger:auto:hold:sensitivity
Syntax:
:trigger:auto:hold:sensitivity?
:trigger:auto:hold:sensitivity
<value>
Function:
The command queries and sets the sensitivity of auto trigger delay.
Parameters:
Name
Tyoe
Range
Default
<value>
discrete
{0|1|2|3}
1
Explanations:
The Table 2- 10 indicates the parameters and their meanings.