Rigol DM3051 Dokumentacja Strona 46

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 123
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 45
RIGOL
© 2008 RIGOL Technologies, Inc.
Programming Guide for DM3000 Series
2-34
Table 2- 7 The parameters and the ranges
Parameter
Range
0
2nF
1
20nF
2
200nF
3
2uF
4
20uF
5
200uF
MIN
2nF
MAX
200uF
DEF
200nF
Example:
To set the capacitance range as 2uF:
:measure: capacitance3
53. :measure:capacitance:range?
Syntax:
:measure:capacitance:range?
Function:
The command queries the measure range of capacitance.
Return Format:
The query returns the current range parameter: 01234 or 5.(see Table 2- 7)
54. :measure:capacitance:digit
Syntax:
:measure:capacitance:digit?
:measure:capacitance:digit{INC|DEC|5|6|7}
Function:
The command queries and sets the digit of capacitance.
Explanations:
Przeglądanie stron 45