RIGOL
© 2007 RIGOL Technologies, Inc.
User’s Guide for DM3000 Series
5-15
Frequency Period Characteristics
Accuracy Specifications (% of reading)
[1]
Function Range
Frequency
Range
1 Year
23±5
Temperature
Coefficient
0 to 18
28 to 55
3Hz-5Hz 0.10 0.005
5Hz-10Hz 0.07 0.005
10Hz-40Hz 0.02 0.005
200mV
to 750V
[2]
40Hz-300kHz 0.02 0.005
3Hz-5Hz 0.10 0.005
5Hz-10Hz 0.07 0.005
Frequency
Period
20mA to10A
[3]
10Hz-10kHz 0.02 0.005
[1] Specifications are for 60 minute warm–up.
[2] For AC input voltages 10% to 120% of range except where noted. 750V range limited to 750VRMS.
100mV range specifications are for full scale or greater inputs. For inputs from 10mV to 100mV, multiply
total % of reading error by 10.
[3] For the 20mA, 200mA, 10A ranges, the AC input current from 10% to 120% of range except where noted.
For 1A range, the AC input current from 50% to 120% of range except where noted.
Measurement Considerations
All frequency counters are susceptible to error when measuring low–voltage, low–frequency signals. Shielding
inputs from external noise pickup is critical for minimizing measurement errors.
Settling Considerations
Errors will occur when attempting to measure the frequency or period of an input following a DC offset voltage
change. The input blocking RC time constant must be allowed to fully settle (up to 1 sec) before the most
accurate measurements are possible.