Rigol DM3064 Podręcznik Użytkownika Strona 143

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 142
RIGOL
© 2007 RIGOL Technologies, Inc.
User’s Guide for DM3000 Series
5-9
Inspection Function on Rear Panel (DM3064)
Including the models which have LAN/GPIB interface, the voltage limit that the LO terminal
relatives to the safety ground is 200Vpeak (Max).
Channels: 12 voltage channels, 4 current channels, appending protection.
Measurement Functions: 2WR, 4WR, Capacitance, DCV, DCI, ACV, ACI, Continuity, Diodes and Frequency.
Work characteristic: The accuracy is the same as measurement on the front panel. The switch time is 50ms
(the inspection plate and the Front-Rear panel) and 600ms (auto range, 4 1/2).
Input characteristic: Diff input voltage 150Vpeak(Max), isolated voltage between channels 150Vpeak(Max),
current input terminal 1Apeak(Max), channel isolation >60dB(@10KHz), all terminal to the
safety ground voltage 150Vpeak(Max).
Input Protection: Inner current channel 2A self- recover fuse, 250V voltage limitation.
Inner voltage channel Expansion cards 500mA fuse.
Real-time Clock
Precision: 1min/month (Environment Temperature >0 )
Clock battery Life: 2 years
Math Functions
Null, Min/Max/Average, dBm, dB, Limit Test (with TTL output)
Other Functions
Reading Hold, Ratio Measurement
High-speed Sampling
50kSa/s (In DataLog function)
Reading Resolution
2,400,000 Count, >6 1/2 digits
USB I/O Interface
USB Host, USB Device interface. It can support U-disk and USB printer.
Other I/O Interface
RS-232, GPIB (Selectable) support for SCPI command, LAN (Selectable)