1-12
Probe Compensation
When the probes are used for the first time, you should compensate the probes to
match the input channels of the oscilloscope. Non-compensated or poorly
compensated probes may cause measurement inaccuracy and error. The probe
compensation procedure is as follows:
1. Perform step 1, 2, 3 and 4 of “Function Inspection” in the previous section.
2. Check the displayed waveforms and compare them with the following figures.
Figure 1-10 Probe Compensation
3. Use a nonmetallic driver to adjust the variable capacitor on the probe until the
displayed waveform is as the Correctly compensated in the figure above.
Over compensated Correctly compensated Under compensated