RIGOL Chapter 3 Performance Test
Service Guide for DS1000E, DS1000D Series
3-12
1MΩ±2%, in parallel with
15pF± 3pF
1MΩ±2%, in parallel with
15pF± 3pF
Voltage accuracy
Relative error between the
average and the actual:
CH1, CH2
2mV, 5mV scale
>5mV scale
± 4%
± 3%
DC linearity
DC voltage error array:
CH1, CH2
200mV scale
5V scale
-2.602.60(mV)
-2.602.60(mV)
DC offset
Relative error of the voltage:
CH1, CH2, 5V scale
06%
Horizontal time
Accuracy
delta-t error
-5.00e-95.00e-9(s)
Trigger sensitivity
Trigger sensitivity error:
10MHz
Full bandwidth
0.11.0div
0.11.5div
Timebase linearity
Period error
0.05.0e-10(s)
Channel delay
Delay:
CH1, CH2
500ps
Trigger delay
Delay:
CH1, CH2
500ps