Chapter 3 Performance Test RIGOL
Service Guide for DS1000E, DS1000D Series
3-9
Logic Analyzer Interface Test (Only for DS1000D Series)
Purpose:
Test if the Logic Analyzer (hereinafter referred to as LA) interface works normally
through the logic signal output module (DG-POD-A) of DG3000.
Tools:
A set of DS1000D series digital oscilloscope
A set of DG3000 Function/ Arbitrary waveform Generator
An active logic head
Two logic cables
17 logic testing leads
A DG-POD-A module
Steps:
1. Connect LA interface on DS1000D with the active logic head by logic cable.
2. Connect DIGITAL OUTPUT interface on the rear panel of DG3000 with
DG-POD-A module by logic cable.
3. Connect the active logic head with DG-POD-A module by logic testing leads, so as
to realize DS1000D and DG3000 connection.
Figure 3-11 DS1000D LA interface connection
4. Power on DS1000D and DG3000 respectively.
5. Turn on DG3000 and press UtilityOutput SetupDigit-ModuPower on to
enable the digital module power on.
6. Press ArbEdit Digital W, select ProtocolPO, Code Pat 32PRBS and adjust
Output Length as 16 Byte as follows, then press Arb button to output digital
signal.
DIGITAL OUTPUT
Interface
DG3000
DS1000D
17 Logic Testing Leads
Logic Cable