Duration Trigger (Only for DS1000D Series)
Duration Trigger Mode: Trigger in appointed time when code terms are satisfied.
Press Trigger MENU→Mode→Duration to display the duration trigger menu as
follows.
Figure 2- 61 Table 2- 37 The Duration Trigger Menu (Page 1/2)
Choose digital channel for Duration Trigger
Figure 2- 62 Table 2- 38 The Duration Trigger Menu (Page 2/2)
Set duration and limit symbol time
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform
and then stop
To go to set up menu, see table 2-40
Key Points:
1. H (High): Logic high: voltage is higher than threshold setting.
2. L (Low): Logic low: voltage is lower than threshold setting.
3. X (Ignore): Don’t care. If all the channels are ignored, the oscilloscope won’t
be triggered.
4. Qualifier: A timer begins when code terms are satisfied. Duration trigger
occurred in the time set by the qualifier.