Chapter 5 To Trigger the Oscilloscope RIGOL
MSO2000A/DS2000A Users Guide 5-19
: trigger on the falling edge of the input signal when the voltage level meets
the specified trigger level.
Idle Time:
Press Idle to set the idle time before the edge counting in Nth edge trigger. The
range available is from 16 ns to 4 s.
Edge Number:
Press Edge to set the value of “N” in Nth edge trigger and the range available is from
1 to 65535.
Trigger Mode:
Press Sweep and select Auto”, “Normal” or ”Single”. For the details, please refer to
Trigger Mode. The corresponding status light of the current trigger mode turns
on.
Trigger Setting:
Press Setting to set the trigger parameters (noise rejection) under this trigger type.
Note: Noise rejection is grayed out and disabled automatically when the signal
source is set to any channel of D0-D15.
Trigger Level:
Use TRIGGER LEVEL to modify the level. For the details, please refer to the
introduction of
Trigger Level.